Perform the Z-port Error Corrections

1. Connect adapter A3 to adapter A2 on port 2. (See Figure 5-12.)

N E T W O R K ANAL’IZER

REFERENCEREFERENCE

PORT 1

pg647e

Figure 5-12. ‘lko-Port Cal Set 1

2.Perform the 2-port error correction using calibration standards appropriate for the connector type at port 1.

Note When using adapter removal calibration, you must save calibration sets to the internal disk, not to internal memory.

Caution Do not mistake the Iine switch for the disk eject button. See the figure below. If the Iine switch is mistakenly pushed, the instrument wiII be turned off, losing all settings and data that have not been saved.

D I S K E J E C T

B U T T O N 1

L I N E SWITCH/

3. Save the results to disk. Name the file “PORTl.”

5 4 2 Optimizing Measurement Results