Page 527
Image 527

kRSPti#SE

&+&p& (Specify Class)

..i :. . . . . .

. When in the specify class more menu, IB!$PSMSE is used to enter the standard numbers for a response calibration. This calibration corrects for frequency response in either reflection or transmission measurements, depending on the parameter being measured when a calibration is performed. (For default kits, the standard is either the open or short for reflection measurements, or the thru for transmission measurements.)

nWhen in the response cal menu, RE$P.C#SE leads to the frequency response calibration. This is the simplest and fastest accuracy enhancement procedure, but should be used when extreme accuracy is not required. It effectively removes the frequency response errors of the test setup for reflection or transmission measurements.

nWhen in the specify class more menu, KESP!BX$ t ‘X%X’1 is used to enter the standard numbers for a response and isolation calibration. This calibration corrects for frequency response and directivity in reflection measurements,

or frequency response and isolation in transmission measurements

nWhen in the response and isolation menu,

~@ZSPf8!&E;:$ J&X?l leads to the menus used to perform a response and isolation measurement calibration, for measurement of devices with wide dynamic range. This procedure effectively removes the same frequency response errors as the response calibration. In addition, it effectively removes the isolation (crosstalk) error in transmission measurements or the directivity error in reflection measurements. As well as the devices required for a simple response calibration, an isolation standard is required.

The standard normally used to correct for isolation is an impedance-matched load (usually 50 or 75 ohms). Response and directivity calibration procedures for reflection and transmission measurements are provided in the following pages.

turns off the tabular listing and returns the measurement display to the screen.

eliminates the need to restart a calibration sequence that was interrupted to access some other menu. This softkey goes back to the point where the calibration sequence was interrupted.

measures the reverse isolation of the calibration standard.

lets you enter a label for the reverse match class. The label appears during a calibration that uses this class.

specifies which standards are in the reverse match class in the calibration kit.

is used to enter the standard numbers for the reverse match (thru) calibration. (For default kits, this is the thru.)

9-36 Key Definitions