NETWORK ANALYZER

I TEST POPT

CABLES

SHORT OPEN SHORT OPEN

FOR S, 1 RESPONSE

pq610e

Figure 5-1.

Standard Connections for a Response Error-Correction for Reflection Measurement

qTo measure the standard when the displayed trace has settled, press:

If the calibration kit you selected has a choice between male and female calibration standards, remember to select the sex that applies to the test port and not the standard.

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement. The analyzer underlines the softkey that you selected after it llnishes the measurement, and computes the error coefficients.

Note This calibration allows only one standard to be measured. If you press the wrong key for a standard, start over with step 6. Do not use a thru standard for a reflection response correction.

Note You can save or store the measurement correction to use for later measurements, that use the same measurement parameters. Refer to the “Printing, Plotting, and Saving Measurement Results” chapter for procedures.

qThis completes the response correction for reflection measurements You can connect and measure your device under test.