Page 596
Image 596

FUNCTIONNOTAVAILABLE

Error Number The function you requested over HP-IB is not avaiiable on the current

202 instrument.

FUNCTIONNOTVALID

Error Number The function you requested is incompatible with the current instrument state. 14

FUNCTIONNOTVALIDDURINGMOD SEQUENCE

Error Number You cannot perform sequencing operations while a sequence is being modified. 131

FUNCTIONNOTVALIDFORINTERNALMEMORY

Error Number The function you selected only works with disk Ivies. 201

FUNCTION ONLYVALIDDURINGMOD SEQUENCE

:~ ..~:::~~:.. .:::. .:., .,~~,~::::::::. ................,.,...,.. ;; ..,,,_

Error Number You can only use the ~~~~::~~~~~~~~...~:.........~,. .. .,... .... ,,.,.,,// .,/ ,, y capability when you are building a

164sequence. Attempting to use this softkey at any other time returns an error message and no action is taken.

HP 8753 SOURCEPARAMETERSCHANGED

Error Number Some of the stimulus parameters of the instrument state have been changed,

61because you have turned correction on. A calibration set for the current measurement parameter was found and activated. The instrument state was updated to match the stimuIus parameters of the calibration state.

This message also appears when you have turned on harmonic mode or frequency offset, and the present frequency range cannot be used with one of these modes.

HPIBCOPY INPROGRESS, ABORTWITHLOCAL

Error Number An HP-IB copy was already in progress when you requested the HP-IB for

169another function. ‘Ib abort the first copy, press ILocal), otherwise the HP-IB is unavailable until the first copy is completed.

lo-l.2 Error Messsges