‘Ihble 7-12. HP 8753E Instrument Specifhtions (4 of 6)

INPUT GENBBAL

DescriptionSpeciftcation

bfAGNITUDE CHABACTBBISTICS

 

Display Resolution

0.01 dB/division

Marker* Resolution

0.001 dB

Dynamic-

IF BW: 10 Hz, averaging factor: 1, inputs: test port 1 and 2 (R to -35 dBm)

For test port powers > -50 dBm, magnitude dynamic accuracy is

0.02dB + 0.001 dB/dB from the reference power (plus the effects of sampler compression).

The following graphs include the effects of noise.

co&

s-: s: s-l

HP8753E Magnitude Dynamic Accuracy 0.3 to 3000 MHz

HP8753E Magnitude Dynamic Accuracy 3-8 GHz

0.01

0.01

10 0 -10 -20 -30 -40 -50 -60 -70 - 8 0 - 9 0 -100

10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 -100

Test Port Power (dBm)

Test Port Power (dBm)

Trace Noiset

 

 

 

3OOkHztoSGHz

<0.006 dB rms

s-1

3GHzt.06GHz

<O.OlO dB rms

s-l

Reference Level

 

 

 

 

f500 dJ3

53

Resoltiion

0.001 dB

s-3

Stability

 

 

 

3OOkHzto3GHz

0.02 dBPC

‘I

3GHzto6GIIz

0.04 dB/w

T

?HAsE cEABAcTEBIsTIcs

 

 

 

 

fl&?O

o

s-3

Display Resolution

0.01

oldivision

s-3

Marker Resolution*

0.01 o

s3

’ Marker resolution for magnitude, phase, and delay is dependent upon the value measured, resolution is limited n5digit8.

. CW sweep, + 6 dBm into ‘l&t Port, ratio measurement, 3 kHz BW

Specifications and Measurement Uncertainties 7-13

Page 466
Image 466
HP manual ‘Ihble 7-12. HP 8753E Instrument Specifhtions 4