HP 8763E (76Q) with Type-N Test Ports

The following specifications describe the system performance of the HP 8753E network analyzer. The system hardware includes the following:

Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . , . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .075

Calibration kit-. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 85036B Cables: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 11857B

lhbie 7-6.

Measurement Port Characteristics (Corrected)* for HP 8753E (750hms) with Type-N ‘Ikst Ports

 

 

 

Frequency Range

 

 

 

 

 

 

 

 

 

 

30 kEz to 300 kEzt

300 kEz to 1.3 GHz

1.3 GHz to 3 GHz

 

Directivity

48 dB

48 dB

43 dB

 

Source match

47 dB

41 dB

36 dB

 

Load match

48 dB

48 dB

43 dB

 

Reflection tracking

60.004 dB

fO.O1O dB

f0.019 dB

 

Transmission tracking

fO.018 dB

9~0.016 dB

f0.033 dB

 

 

 

 

 

 

*Applies at 26 f6 ‘C t Typical Performance

HP8753E With HP85036B Calibration Kit

HP8753E With HP85036B Calibration Kit

 

 

 

 

0 1

 

-30

-50

-70

-10

-30

-50

-70

10

-10

S21 Transmission

Coefftcient (dB)

 

S21 Transmission

Coeffkient (dB)

 

HP8753E Wiih HP85036B Calibration Kit

HP8753E Wlth HP85038B Calibration Kit

 

 

 

8.0

 

 

 

 

 

 

0.2

0 4

0.6

0.8

0.2

0.4

0.6

0.8

1.C

Sl 1 Reflection Coefficient

(linear)

Sl 1 Reflection

Coeffkient (linear)

 

7-6 Specifications and Measurrwmt Uncertainties

Page 459
Image 459
HP 8753E manual HP 8763E 76Q with Type-N Test Ports, GHz to 3 GHz