Figure 2-42. Example Flat Limit Lines

Creating a Sloping Limit Line

This example procedure shows you how to make limits that test the shape factor of a SAW titer. The following limits are set:

Frequency Bange ................................................................. Power Range

123MHzto125MHz.. ........................................................ -65dBto-26dB

144 MHz to 146 MHz

-26dB to -65 dEI

1. To access the limits menu and activate the limit lines, press:

2 . To establish the start frequency and limits for a slopinglimit line that tests the low side of the filter, press:

Making Measurements 249

Page 100
Image 100
HP 8753E manual Creating a Sloping Limit Line, Example Flat Limit Lines