Verification Kit

Accurate operation of the analyzer system can be verified by measuring known devices other than the standards used in calibration, and comparing the results with recorded data.

HP 86029B 7-mm Verifkation Kit

This kit contains traceable precision 7-mm devices used to confIrm the system’s error-corrected measurement uncertainty performance. Also included is verification data on a 3.5 inch disk, together with a hard-copy listing. A system verification procedure is provided with this kit and alsointhe HP8753EServiceGuide.

Test Port Return Cables

The foIIowing RF cables are used to connect a two-port device between the test ports. These cables provide shielding for high dynamic range measurements.

HP 11867D 7-mm l&t Port Return Cable Set

This set consists of a pair of test port return cables that can be used in measurements of 7-mm devices. They can also be used with connectors other than 7 mm by using the appropriate precision adapters.

HP 11867B 75 Ohm Type-N Test Port Return Cable Set

This set consists of test port return cables for use with the HP 8753E Option 075.

Adapter Kits

HP 11862B 50 to 75 Ohm Minimum Loss Pad.

This device converts impedance from 50 ohms to 75 ohms or from 75 ohms to 50 ohms It is used to provide a low SWR impedance match between a 75 ohm device under test and the HP 8753E network analyzer (without Option 075).

These adapter kits contain the connection hardware required for making measurements on devices of the indicated connector type.

nHP 11853A 50 Ohm Type-N Adapter Kit

nHP 11854A 50 Ohm BNC Adapter Kit

nHP 11855A 75 Ohm Type-N Adapter Kit w HP 11856A 75 Ohm BNC Adapter Kit

1 l-2 Compatible Peripherals

Page 619
Image 619
HP 8753E manual Verification Kit, Test Port Return Cables, Adapter Kits