MEMORYFORCURRENTSEQUENCEISFULL

Error Number All the memory in the sequence you are modifying is filled with instrument

132 commands.

MORESLIDESNEEDED

Error Number When you use a sliding load (in a user-dellned calibration kit), you must set at

71 least three slide positions to complete the calibration.

NOCALIBRATIONCURRENTLYINPROGRESS

Error Number The ~~~~~~~~~~~softkey is not valid unless a calibration is *eady

.I .,,... ::.~.~~~:~:~~.~:~~....~,...,.,;;;..~......~.~~:~:~~~~:~:~~~~

.._

,..:,.......__

6 9in progress. Start a new calibration.

NODISKMEDIUMINDRIVE

Error Number You have no disk in the current disk unit. Insert a disk, or check the disk unit

41 number stored in the analyzer.

NOFAIL FOUND

Service Error The self-diagnose function of the instrument operates on an internal test Number 114 failure. At this time, no failure has been detected.

NOFILE FOUND ONDISK

Error Number No flies of the type created by an analyzer store operation were found on the

4 5disk or the disk drive is empty. If you requested a specific file title, that llle was not found on the disk.

NO IF FOUND: CHECKR INPUTLEVEL

Error Number The first IF signal was not detected during pretune. Check the front panel R

5channel jumper. If there is no visible problem with the jumper, refer to the HP 8753E Network Anulgzer semrice Guide for troubleshooting.

Error Messages lo-16

Page 599
Image 599
HP 8753E manual 9in progress. Start a new calibration