SEGMENT#nPOUEROUTSIDERANGELIMIT

Information The selected power range does not support the power level of one or more

Message segments in the swept list table. This message appears when swept list mode is not on and reports the llrst segment that is out of range. Change the segment power or change the power range.

SEGMENT#nSTARTFREQ OVERLAPSPREVIOUSSEGMENT

Information A segment entered in the swept list table caused one or more frequency

Message segments to overlap. This message appears when swept list mode is not on and reports the first segment that is overlapping another. Change the frequency ranges of the overlapping segments.

SELECTEDSEQUENCEISENPTY

Error Number The sequence you attempted to run does not contain instnunent commands 124

SELFTESTkFAILED

Service Error Internal test #n has failed. Several internal test routines are executed at Number 112 instrument preset. The analyzer reports the iirst failure detected. Refer to the

HP 8753E NetworkAmlgm Semrice G&de for troubleshooting information on internal tests and the self-diagnose feature.

SEQUENCE ABORTED

Error Number The sequence running was stopped prematurely when you pressed the llocalJ

157 key.

SEQUENCEMAYHAVECHANGED,CAN'TCONTINUE

Error Number When you pause a sequence, you cannot continue it if you have modified it.

153 You must start the sequence again.

Error Messages 1O-23