Frequency Response and Isolation Error-Corrections

nremoves frequency response of the test setup

nremoves isolation in transmission measurements w removes directivity in reflection measurements

You can make a response and isolation correction for the following measurements:

w reflection measurements

N transmission measurements

ncombined reflection and transmission measurements

Response and Isolation Error-Correction for Reflection Measurements

Although you can perform a response and isolation correction for reflection measurements, Hewlett-Packard recommends that you perform an S1l one-port error-correction; it is more accurate and just as convenient.

1.Press B.

2.Select the type of measurement you want to make.

•I If you want to make a reflection measurement on PORT 1 (in the forward direction, &I), leave the instrument default setting.

qIf you want to make a reflection measurement on PORT 2 (in the reverse direction, S&, press:

3.Set any other measurement parameters that you want for the device measurement: power, sweep type, number of points, IF bandwidth.

4.To access the measurement correction menus, press:

Ical]

.,:_.,. .,...;:.:...,...:.:.:.:.:.:.:.:

,,

,.I

__/, .,..,..,.,. _

6.If your calibration kit is different than the kit specified under the ~~~~~~~~~~~~~;~ softkey,

i..ii..-.-....

press:

...... i...........T..................... : _y: .,

~~~~~~,~~~,~ ~~~~~~~~~~~~ (select you type of kit) ;~~

,,,

,,,

,, ,,,

,,,..,.

* _ _

..-....

/

.._

 

 

.._

 

If your type of calibration kit is not listed in the displayed menu, refer to the “Modifying Calibration Kit Standards” procedure, located later in this chapter.

6.To select a response and isolation correction, press:

7.Connect the short or open calibration standard to the port you selected for the test port (PORT 1 for S11 or PORT 2 for $2).

Note

Include any adapters that you will have in the device measurement. That is,

 

connect the standard device to the particular connector where you will connect

 

your device under test.

 

 

5-14 Optimizing Measurement Results