Restarting a Calibration

If you interrupt a calibration to go to another menu, such as averaging, you can continue the

ii .,,.

c&brationby pressing& ~,~~.~.~~~~~ softkey in the co~e&ionmenu.

Cal Kit Menu

The cal kit menu provides access to a series of menus used to specify the characteristics of calibration standards. The following softkeys are located within the cal kit menu:

The Select Cal Kit Menu

,,;.? : "'i. ,,/

fiessbg the ~~~~~~~~~~~~ softkey with the cd kit menu provides access to the select

cal kit menu. This menu allows you to select from several default calibration kits that have different connector types These kits have predeiined standards and are valid for most applications. It is not possible to overwrite these standard de&ritions

The numerical definitions for most Hewlett-Packard calibration kits can be referenced in the calibration kit operating and service manuals, or can be viewed on the analyzer. The standard deiinitions can also be modified to any set of standards used.

642 Application and Operation Concepts