HP 8753E manual Application and Operation Concepts

Models: 8753E

1 699
Download 699 pages 49.83 Kb
Page 403
Image 403

Receiver and Source Requirements. Refer to Chapter 7, “Specifications and Measurement Uncertainties. n

IF Input: R always; and port 1 or port 2 for a ratio measurement.

Display Annotations. The analyzer shows the annotation of s when the frequency offset mode is on. The annotation of? indicates that the source frequency is approximately 10 MHz away from the sum of the IF and Lo frequencies that you requested. This is most likely caused by the Lo frequency being outside the - 1 to + 5 MHz accuracy requirement.

Error Message. If you connect your test device before you switch on the frequency offset function, the error message PHASE LOCK CAL FAILED appears on the screen. This is normal, and

:

fl (-J&appear when you press the ~~il~~~~,~~~~~~~~~ s-&&y. ii ..s . . . . ...i....... i.;.;; ii. ..ws. .................:A...::.<..

Spurious Signal Passband Frequencies. Unwanted mixing products (or spurious Lo signals) at specific frequencies can cause measurement inaccuracy, because of the characteristics of a sampler. These specific frequencies can be calculated. You can reduce unwanted mixing products going to the sampler by inserting a low pass filter at your test device’s IF output.

6-122 Application and Operation Concepts

Page 403
Image 403
HP 8753E manual Application and Operation Concepts