Frequency Response Error-Corrections

You can remove the frequency response of the test setup for the following measurements:

nreflection measurements

w transmission measurements

ncombined reflection and transmission measurements

Response Error-Correction for Reflection Measurements

1.Press w).

2.Select the type of measurement you want to make.

q If you want to make a reflection measurement on PORT 1 (in the forward direction, Sil), leave the instrument default setting.

q If you want to make a reflection measurement on PORT 2 (in the reverse direction S&, press:

qSet any other measurement parameters that you want for the device measurement: power, sweep type, number of points, or IF bandwidth.

qTo access the measurement error-correction menus, press:

Lcal]

0 If your calibration kit is different than the kit speci~ed under the ::~~~~~~~:~~~~, softkey,

press:

If your type of calibration kit is not listed in the displayed menu, refer to the “Modifying Calibration Standards” procedure, located later in this chapter.

0 To select a response correction, press:

Icall ~~~~~~~ ~~~~~~~

.................... .................................- ......... .:

qConnect the short or open calibration standard to the port you selected for the test port (PORT 1 for S11 or PORT 2 for !&).

Note Include any adapters or cables that you will have in the device measurement. That is, connect the standard device to the particular connector where you will connect your device under test.

Optimizing Measurement Results 54