PLOT ABORTED

Error Number When you press the LLocal) key, the analyzer aborts the plot in progress. 27

PLOTTER:noton,notconnect, wrongaddrs

Error Number The plotter does not respond to control. Verify power to the plotter, and check

26the HP-IB connection between the analyzer and the plotter. Ensure that the plotter address recognized by the analyzer matches the HP-IB address set on the plotter itself.

PLOTTERNOTREADY-PINCHWHEELSUP

Error Number The plotter pinch wheels clamp the paper in place. If you raise the pinch

28 wheels, the plotter indicates a “not ready” status on the bus

POSSIBLEFALSELOCK

Error Number Phase lock has been achieved, but the source may be phase locked to the

6wrong harmonic of the synthesizer. Perform the source pretune correction routine documented in the “Adjustments and Correction Constants” chapter in the HP 875333 Network Analyzer semrice Guide.

POWERMETERINVALID

Error Number The power meter indicates an out-of-range condition. Check the test setup. 116

POWERMETERNOT SETTLED

Error Number Sequential power meter readings are not consistent. Verify that the equipment

118 is set up correctly. If so, preset the instrument and restart the operation.

POWERSUPPLYHOT!

Error Number The temperature sensors on the A8 post-regulator assembly have detected an

21over-temperature condition. The power supplies regulated on the post-regulator have been shut down.

Error Messages 10-18

Page 603
Image 603
HP 8753E manual Plot Aborted