nAccuracy

qAccuracy enhancement methods that range from normalizing data to complete one or two port vector error correction with up to 1601 measurement points, and TRL*/LRM*. (Vector error correction reduces the effects of system directivity, frequency response, source and load match, and crosstalk.)

qPower meter calibration that allows you to use an HP-IB compatible power meter to monitor and correct the analyzer’s output power at each data point. (The analyzer stores a power correction table that contains the correction values.)

w Printin& Plotting, and Saving

qDirect print or plot output of displayed measurement results, with a time stamp if desired, to a compatible peripheral with a serial, parallel, or HP-IB interface.

qInstrument states storage in internal memory for the following times, or on disk indefinitely.

Temperature at 70 “C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250 days (0.68 year) characteristically Temperature at 40 OC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1244 days (3.4 years) characteristically Temperature at 25 “C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 years characteristically

qLIF/DOS disk formats for saving instrument states and measurement data.

qIntegration of a high capacity micro-floppy disk drive.