Response Error-Correction for Transmission Measurements

1.Press LPresetJ.

2.Select the type of measurement you want to make.

qIf you want to make a transmission measurement in the forward direction (&), press:

qIf you want to make a transmission measurement in the reverse direction (&a), press:

3.Set any other measurement parameters that you want for the device measurement: power, number of points, IF bandwidth.

4.To select a response correction, press:

5.Make a “thru” connection between the points where you will connect your device under test.

Note Include any adapters or cables that you will have in the device measurement. That is, connect the standard device where you will connect your device under test.

NETWORK ANALYZER

T E S T P O R T

P O S S I B L E

A D A P T E R S

pg611e

Figure 5-2.

Standard Connections for Response Error-Correction for Transmission Measurements

6.‘Ib measure the standard, press:

;,.../___ _

;2J. $!B&

The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement.

The analyzer underlines the ~~ softkey after it measures the calibration standard, and - .._...._........

computes the error coefficients.

Optimizinfl Measurement Results 5-l 1

Page 233
Image 233
HP 8753E manual Response Error-Correction for Transmission Measurements, 2J. $!B