NETWORK ANALYZER

F O R S,, R E S P O N S E

S H O R TO P E N

F O RS,, ,SOLATlON b

L O A DL O A D

F O R S22 R E S P O N S E

pg612e

Figure 5-4.

Standard Connections for a Response and Isolation Error-Correction for Reflection

Measurements

8.To measure the standard, press:

If the calibration kit you selected has a choice between male and female calibration standards, remember to select the sex that applies to the test port and not the standard.

The analyzer displays WAIT - MFASURING CAL STANDARD during the standard measurement. The analyzer underlines the softkey that you selected after it finishes the measurement, and computes the error coefficients

9.Connect the load calibration standard to the test port.

10.lb measure the standard for the isolation portion of the correction, press:

11.To compute the response and directivity error coefficients, press:

The analyzer displays the corrected SH (or S& data. The analyzer also shows the notation Cor to the left of the screen, indicating that the correction is switched on for this channel.

Note

You can save or store the error-correction to use for later measurements.

 

Refer to the “Printing, Plotting, and Saving Measurement Results” chapter for

 

procedures.

 

 

12.This completes the response and isolation error-correction for reflection measurements. You can connect and measure your device under test.

Optimizing Measurement Results 5-15