Modify the Cal Kit Thru Deilnition

With this method it is only necessary to use adapter B. The calibration kit thru definition is modified to compensate for the adapter and then saved as a user kit. However, the electrical delay of the adapter must hrst be found.

1.Perform a l-port calibration on PORT 2.

2.Connect adapter B to the test port.

3.Add a short to the open end of the B adapter.

4.Measure the delay of the adapter by pressing (G) %&AY.

5.Divide the resulting delay measurement by 2.

6.Determine the offset delay of the calibration short by examming the dehne standard menu (see “Dehne Standard Menus”).

7.Subtract the short offset delay from the value calculated in step 5. This corresponds to the delay of adapter B.

8.Modify the calibration kit thru de8nition by entering in the electrical delay of adapter B. Save this as a user kit.

9.Perform the desired calibration with this new user kit.

10.Measure the test device.