3.lb activate a marker to measure the group delay at a particular frequency, press: (%ZQ and turn the front panel knob, or enter a value from the front panel keypad.

Figure 2-37. Group Delay Example Measurement

Group delay measurements may require a specific aperture (AF’) or frequency spacing between measurement points The phase shift between two adjacent frequency points must be less than 180°, otherwise incorrect group delay information may result.

4. ‘lb vary the effective group delay aperture from minimum aperture (no smoothing) to appro~ately 1% of the frequency span, press: LAvg) ~~~~~~~~~~~~

:::::. .:::..:::..:..; ....A..s.;..i. ... .2......22;;;: ......... ...A.:.:::*

When you increase the aperture, the analyzer removes fine grain variations from the response. It is critical that you specify the group delay aperture when you compare group delay measurements.

aw000008

Figure 2-38. Group Delay Example Measurement with Smoothing

244 Making Measurements

Page 95
Image 95