Amplifier Tksting

Amplifier parameters

The HP 8753E allows you to measure the transmission and reflection characteristics of many amplifiers and active devices. You can measure scalar parameters such as gain, gain flatness, gain compression, reverse isolation, return loss (SWR), and gain drift versus time. Additionally, you can measure vector parameters such as deviation from linear phase, group delay, complex impedance and AM-to-PM conversion.

pg6137d

Figure 6-82. Amplifier Parameters

The analyzer allows you to make a swept-frequency measurement of an amplifier’s second or third harmonic as shown in F’igure 6-83.

 

 

 

 

 

 

 

 

21 Jun

1 9 9 4 12:41. 19

Cl-l1

B&M

log

MAG

10

dB/

REF

0

dB

1-25.744

dB

CH2

B

log

MAG

10

dB/

REF

0

dB

I-51.569

dB

PRm

H=2

PRm

H=3

 

 

 

 

 

 

 

 

 

 

CHl

START

2 0 . 0 0 0

000

MHz

STOP 1

000

000

000

MHz

pg6239

CH2

START

2 0 . 0 0 0

000

MHz

STOP 1

000

000

000

MHz

 

Figure 6-83.

Swept Frequency Amplifier Measurement of Absolute Fundamental, 2nd and 3rd

Earmonic Output Levels

Application and Operation Concepts 6.163