ASCII: MISSING ‘VAR’ STATEMENT

Error Number The CITblle you just downloaded over the HP-IB or via disk was not properly

196 organized. The analyzer is unable to read the “VAR” statement.

AVERAGING INVALID ON NON-RATIO MEASURE

Error Number You cannot use sweep-to-sweep averaging in single-input measurements.

13Sweep-sweep averaging is valid only for ratioed measurements (A/R, B/R, A/B, and S-parameters). You can use noise reduction techniques, such as narrower IF bandwidth, for single input measurements

BAD FREQ FOR HARMONIC OR FREQ OFFSET

Error Number You turned on time domain or recalled a calibration that resulted in start and

181 stop frequencies that are beyond the allowable limits.

BATTERY FAILED. STATE MEMORY CLEARED

Error Number The battery protection of the non-volatile CMOS memory has failed. The CMOS

183memory has been cleared. Refer to the HP 8753E Network Analgzer semrice Guide for battery replacement instructions See Chapter 12, “Preset State and Memory Allocation,” for more information about the CMOS memory.

BATTERYLOW! STORE SAVE REGS TO DISK

Error Number The battery protection of the non-volatile CMOS memory is in danger of failing.

184If this occurs, all of the instrument state registers stored in CMOS memory will be lost. Save these states to a disk and refer to the HP 8753E Network Anulgzer Seruicc G&&e for battery replacement instructions. See Chapter 12, “Preset State and Memory Allocation,” for more information about the CMOS memory.

BLOCK INPUT ERROR

Error Number The analyzer did not receive a complete data transmission. This is usually

34caused by an interruption of the bus transaction. Clear by pressing the ILocal key or aborting the I/O process at the controller.

1 O-4 Error Messages

Page 588
Image 588
HP 8753E manual Ascii Missing ‘VAR’ Statement