CAN'TSTORE/LOADSEQUENCE, INSUFFICIENTMEMORY

Error Number Your sequence transfer to or from a disk could not be completed due to

127 insufficient memory.

CAUTION: AUXCHANNELSMEASURES-PARAMETERS ONLY

Error Number This message appears if you try to assign a non-S-parameter measurement to an

216 auxiliary channel.

CAUTION: CORRECTIONOFF: AUXCHANNEL(S) DISABLED

Error Number This message is displayed when correction is forced off due to a stimulus

215change that is not compatible with the current calibration while an auxiliary

channel is enabled. The auxiliary channels are restored when correction is turned on by pres,&,g Lcal] ~~~~~~~~~~~.

CHl (CH2) TARGETVALUENOTFOUND

Error Number Your target value for the marker search function does not exist on the current

159 data trace.

CONTINUOUS SWITCHINGNOTALLOWED

Error Number Your current measurement requires different power ranges on channel 1 and channel 2. To protect the attenuator from undue mechanical wear, test set hold

10will be activated.

The “tsH” (test set hold) indicator in the left margin of the display indicates that the inactive channel has been put in the sweep hold mode.

COPY: devicenotresponding; copy aborted

Error Number The printer or plotter is not accepting data. Verify the cable connections, HP-IB

170 addresses, and otherwise ensure that the copy device is ready.

COPYOUTPUTCOMPLETED

Information The analyzer has completed outputting data to the printer or plotter. The

Message analyzer can now accept another copy command.

1 O-6 Error Messames

Page 590
Image 590
HP 8753E manual CANTSTORE/LOADSEQUENCE, Insufficientmemory