Function check

 

Operation

 

 

 

 

 

 

Check

Result

Remark

 

 

 

7. Suppression (Reject)

 

 

 

 

 

For settings see 2.

 

 

Set the 2nd echo just under 80% SH

 

 

 

 

 

REJECT: SET 80%

Echos 2 to 5 are consecutively suppressed.

2nd echo is visible when the gain

 

Only the initial pulse and the 1st echo at Div. 2

is increased by 0.5-1 dB.

 

are visible.

 

 

 

 

Set REJECT to 0% after check

 

 

 

 

 

8. Vertical linearity

 

 

 

 

 

For settings see 2.

Height of echo is displayed in the

 

GATE LOGIC: POSITIVE

measurement line in % screen HT (H=80%)..

 

a-THRESH: 10%

 

 

a-WIDTH: 50 mm

 

 

AMPLITUDE: % screen HT GAIN and

 

 

a-START so that selected echo is in the

 

 

gate

 

 

 

 

 

Increase GAIN by 2dB, then decrease

Amplitude values:

Deviations must not be more then

by 2 dB

+ 2 dB = 100%

2% screen HT.

 

-2 dB = 63%

 

 

-4 dB = 50%

Determine the height at the last

 

-6 dB = 40%

 

-8 dB = 32%

measured value, if required, by

 

-10 dB = 25%

comparison with the graticule

 

-12 dB = 20%

because values below the gate

 

-14 dB = 16%

are not displayed

 

-16 dB = 13%

 

 

-18 dB = 10%

 

 

 

 

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-67

Page 112
Image 112