Operation

Measuring thickness

 

 

5.7 Measuring thickness

USN 52R/USN 52L enable you to measure thickness and distance in a simple way. The measured value is directly displayed. One of the following two methods of measurement is used, depending on the application:

Measurement to the first echo – e.g. corrosion measurement with a dual (T/R) probe

Measurement of the multi-echo sequence with a single- element probe on a test object which has a good surface on both sides

Preparations

Select probe

Set MEASURE (refer to previous section)

Select TOF (refer to Page 5-9)

Calibrate A-scan display

Calibrate the USN 52 R for the selected probe and the measurement mode (refer to previous section)

Measuring the thickness

Clean the surface of the test object from dirt and any loose particles

Apply couplant to the surface of the test object

Using slight and even pressure, couple the probe

Observe A-scan display and S=...”-value until the measured value display is stable

Maximize the echo amplitude

Read measured value

HNote:

If you are operating with MULTECHO in MEASURE, observe Gate b in order to evaluate the correct echo.

You can store the displayed measurement values in the Data Logger or transfer them into another instrument via the RS232 interface.

5-34

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

Page 79
Image 79
Epson USN 52R, USN 52L manual Measuring thickness, Preparations, Measuring the thickness