Operation

Adjustment of the USN 52R/USN 52L

 

 

Setting the pulser

In the function group PULS you can change the operat- ing mode and the pulse repetition rate of the pulser as well as operate the reject function.

– Use JC to go to the function group PULS.

PULSER Changeover of initial pulse

You can use this function to change between the single- element or dual (T/R) mode for the instrument opera- tion:

SINGLE: single-element operation mode; the probe connector sockets are connected in parallel.

DUAL: dual (T/R) operation mode; the initial pulse is available at the green socket - the red socket is connected to the amplifier input.

Select this setting if you are using a dual (T/R) probe or the through-transmission mode.

Operation:

– Select the required setting in PULSER using M.

HNote:

The corresponding symbol will be displayed in the sta- tus line:

Single Dual

5-8

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

Page 53
Image 53