Operation

Adjustment of the USN 52R/USN 52L

 

 

Calibration functions

USN 52R/USN 52L are accurately calibrated using the functions of the function group S-CAL.

– With I go the function group S-CAL.

HNote:

In order to obtain the correct setting, you must have selected the display range, according to the test task, in the function group BASIC beforehand.

HNote:

Read Chapter 5.3 first in order to exactly set the sound velocity and the probe delay.

HNote:

Refer also to the AUTO CAL function in Chapter 5.3. The function group S-CALcorrespondingly changes when this function is used.

a-START Gate start

The active value for the gate start, which you have entered into the function group GATE, is displayed in a-START.If necessary, you can make a subsequent change to this value here.

5-12

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

Page 57
Image 57