Operation

Adjustment of the USN 52R/USN 52L

 

 

Selection of basic data

In this function you set the different values which serve as a measurement basis and for the display indication.

– With JC go to the BASIC function group.

RANGE Calibration range

In this function you set the display range for the A-Scan. In doing this, you can select between two settings:

coarse: in steps

-5 mm

-10 mm

-25 mm

-50 mm

-100 mm

-125 mm

-200 mm

-250 mm

-500 mm

-1000 mm

-5000 mm

fine: continuous from 5.0 mm bis 5000 mm. The function name appears in small letters (“Range”).

Operation:

Press N or O in order to change between the steps or to continuously select the required value.

Coarse/Fine change-over: Simultaneously press N and Oin RANGE.

5-4

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

Page 49