Operating the DGS

Storing a new probe

For general information on operation, please refer to the USN 52 R Operating Manual.

AAttention:

You cannot estimate any flaw sizes with the DGS op- tion of the USN 52R/USN 52L. You can only compare the reflecting power of a natural flaw with that of a theo- retical flaw (= disk-shaped reflector) at the same depth.

3.1 Storing a new probe

There are 13 probes including the corresponding param- eters stored in the instrument. You can program another probe.

Switch to the second operation level: X

Tag the function table DGS: T or V:

Tag the left column of this function table: J or C

Select no. 16 in the PROBE-#function.

DGS-18

Issue 02, 03/99

Krautkramer USN 52R/USN 52L Option DGS

Page 193