DAC/TCG

Operation

 

 

The two settings DAC and TCG can only be switched on if a curve was recorded beforehand, or if a data set containing a curve was recalled.

REC: Use this setting to prepare the instrument for the recording of a DAC curve. A curve that possibly already exists is deleted.

a-START:With this, you can determine the starting point of the gate (please also refer to function group GATE). This function is repeated here because it is required for recording a curve and for the echo evalua- tion.

DAC ECHO: The DAC ECHO function shows the num- ber of curve points (reference echoes) and is used in the recording mode (REC) for recording the next refer- ence echo.

Operation:

Before starting to record a reference curve, you have to carry out a correct instrument calibration first. This es- pecially applies to the functions: RANGE, MTL VEL, FREQ., RECTIF. and DAMPING due to the fact that a variation of these functions is no longer possible with an active DAC or TCG.

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-27

Page 72
Image 72
Epson USN 52L, USN 52R manual Operation