Configuring the USN 52R/USN 52L

Operation

 

 

GRATICULE Display graticule

In this function you either switch the display graticule on or off. In doing this, you are able to directly read off the echo positions and heights.

Operation:

Highlight the left function group in the left function table with JC.

Enable or disable GRATICULE with M. The currently active setting is displayed.

AMPLITUDE Amplitude display

In this function you determine how the amplitude of the highest echo is to be displayed in the gate. This ap- pears in the measurement line as an “H = ...”-value. You have the following possibilities:

% Screen HT: A percentage of the screen height

dB THRESHOLD: (dB REF not active) expressed as a dB difference relative to the gate threshold

dB THRESHOLD: (dB REF active) expressed as a dB difference relative to the stored reference echo amplitude

Operation:

Highlight the left function group in the left function table with I.

With M select the required setting.

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-43

Page 88
Image 88
Epson USN 52L, USN 52R manual Graticule Display graticule, Amplitude Amplitude display