Storing measured values - Data Logger

Operation

 

 

5.8Storing measured values - Data Logger

The Data Logger helps you to store and manage your measurement results. These are saved even when the batteries are removed. The stored measured values can be:

displayed

deliberately deleted

transferred to a peripheral

output via a PC or printer

A number of measurements are compiled into one

“FILE”. To file measured values in the Data Logger you must

determine the required number of FILES

select a FILE

Determining the number of FILES

You determine the total number of FILES (max. 99) – the instrument automatically determines the number of measurement locations according to the following for- mula:

Number of measurement locations = 2500 : Number of FILES

The first measured value of each FILE is automatically allocated the number 1. However, you can also select another measurement location. All the following mea- sured values are consecutively filed until all measure- ment locations are full or until you select another FILE or measurement location.

HNote:

You can only determine the number of FILES when the Data Logger is empty.

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-35

Page 80
Image 80