Operation

 

Function check

 

 

 

 

 

 

Check

Result

Remark

 

 

 

13. Gate threshold

 

 

 

 

 

For settings see 2.

 

 

GATE LOGIC: POSITIVE

 

 

a-THRESH: 40%

 

 

a-START: 30 mm

 

 

a-WIDTH: 40 mm

 

 

1st Echo to 38% screen HT.

 

 

 

 

 

Increase GAIN setting by 1 dB.

Both LEDs flash alternately when the echo

 

 

exceeds the gate.

 

 

 

 

a-THRESH: 80%

 

 

1st Echo to 78%

 

 

 

 

 

Increase GAIN setting by 1 dB.

Both LEDs flash alternately when the echo

 

 

exceeds the gate.

 

 

 

 

14. Memory

 

 

 

 

 

Set any echo display and store with

The echo display and settings are stored.

 

STORE: ON.

 

 

 

 

 

Remove N30.

Echo display disappears.

 

 

 

 

RECALL: ON.

Echo display appears. Echo display and

 

 

settings must correspond to the condition

 

 

when stored.

 

 

 

 

Switch off instrument Switch on after

Echo display appears again until a parameter

 

about 30 seconds. RECALL: ON.

is changed (e.g. GAIN).

 

 

 

 

5-70

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

Page 115
Image 115
Epson USN 52R, USN 52L manual Gate threshold