Setting the flaw location calculation

Operation

 

 

 

5.12Setting the flaw location calculation

When using angle-beam probes, the flaw location can be automatically calculated. The following values are indicated on the display:

Sound path

Reduced projection distance: The distance from the front edge of the probe to the location of the flaw, projected on the surface

Depth: Distance from flaw location to surface

The illustration on the next page shows the relationship between these values.

HNote:

The gate must be enabled and the MEASURE function set to “0 TO 1st“.

With U go to center function table.

All the necessary functions are available in the right group.

P

projection distance

PR

reduced projection distance

Xdistance probe edge – beam index point

S effective sound path

D distance flaw location – surface

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-59

Page 104
Image 104