Overview of the DGS menu

DGS mode

Menu functions

Function

Description

Ranges/Options

Keys

GAIN

With the DGS mode switched on, the

0.0 to 110.0 dB in steps

B switchover between reference gain

 

GAIN function indicates reference gain

of 0.5, 1.0, 2.0, 6.0 or

and incremental gain

 

and incremental gain.

user defined

N - decrement

 

 

 

O - increment

 

 

 

 

ERS

The ERS function (Equivalent Reflector

INFINITE 0.5 to 35 mm

N - decrement

 

Size) defines the recording level.

limited by probe

O - increment

 

 

parameters

 

 

 

 

 

DGS

The DGS function switches the DGS

ON, OFF

N or O

MODE

mode on or off.

 

 

 

 

 

 

A-START

The A-START function defines the

0.000 to 5000 mm

N - Gate to the lefte t o t h e l e f t

 

horizontal position of the gate.

continuously variable

O - Gate to the right

 

 

 

 

DGS-12

Issue 02, 03/99

Krautkramer USN 52R/USN 52L Option DGS

Page 187
Image 187
Epson USN 52R, USN 52L manual Menu functions, Gain