Operation

 

Function check

 

 

 

 

 

 

 

 

Check

Result

Remark

 

 

 

 

 

9. Damping

 

 

 

 

 

 

 

For settings see 2. Also:

The echo drops in each case to approx.

The difference is approx.

 

DAMPING: 1 K ohms.

40% (with 150 ohms),

-6 dB

 

Set an echo to 80% screen HT, then

30% (with 75 ohms),

-8 dB

 

DAMPING to 150, 75 and 50 ohms.

20% (with 50 ohms).

-12 dB

 

 

 

 

 

10. Resolution test

 

 

 

 

 

 

 

For settings see 2. Also:

There should be no rise in the sweep between

 

 

RANGE: 5 mm

the trailing flank of the delay echo and the

 

 

RECTIF.: NEG HALF

rising flank of the 1st backwall echo.

 

 

Connect CLF4.

 

 

 

Shift the 1st Echo of the delay path to

 

 

 

the left edge of the display using DELAY,

 

 

 

Couple the CLF4 to 1 mm steel.

 

 

 

Adjust the 1st echo from the backwall

 

 

 

echo sequence to 80% screen HT.

 

 

 

 

 

 

 

11. Pulser/Receiver cross-talk

 

 

 

 

 

 

 

For settings see 2.

Read off dB REF value:

Echo must exceed gate, otherwise

 

PULSER: SINGLE

dB-REF value should be 60 dB.

db REF cannot be activated.

 

GATE LOGIC: POSITIVE

 

 

 

a-THRESH: 10%

 

 

 

a-START: 30 mm

 

 

 

a-WIDTH: 50 mm

 

 

 

N30 in Pulser socket

 

 

 

GAIN so that 1st echo is just over 20%

 

 

 

screen HTH.

 

 

 

Press dB REF, PULSER: DUAL

 

 

 

Increase GAIN until the echo or noise is

 

 

 

over 20% screen HT.

 

 

 

 

 

 

 

 

 

 

 

5-68

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

Page 113
Image 113