Adjustment of the USN 52R/USN 52L

Operation

 

 

REJECT Suppression of spurious echoes

REP-RATE Repetition rate

With the REJECT function you are able to suppress unwanted signals, e.g. material noise.

The setting, given in percent, indicates the height that echoes have to reach in order to be displayed in the first place.

Operation:

In REJECT set the required percentage value with M.

AAttention:

Be careful with this function because flaw echoes can also be suppressed.

The number of the ultrasonic signals emitted per sec- ond is varied using this function. This enables selective suppression of spurious or interference echoes and especially phantom echoes which may occur when testing larger test objects. The function enables two settings:

HIGH: Maximum pulse repetition rate

(113 to 905 Hz, depending on the range setting); default setting.

LOW: Reduced pulse repetition rate

(56 to 452 Hz, depending on the range setting); for testing larger test objects or in case any statistical noise signals occur.

Operation:

In REP-RATEset the required percentage value with M.

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-9