Codes for remote control

The USN 52R/USN 52L option DGS can receive coded commands from a PC. To enable this you must connect the instrument with the PC via the RS232 interface.

(Please refer to USN 52R/USN 52L Operating Manual, chap. 8. In addition, you will find in the Operating Manual the codes for the remote control of all instru- ment functions except for the DGS function.)

The tables show the codes for the remote control of the DGS mode, the ranges or parameters, and the resolu- tion.

The functions are arranged in alphabetical order accord- ing to the names appearing in the menu.

The column “Code” shows the letter combinations that you should enter on the PC in order to start the corre- sponding function on the USN 52R/USN 52L option DGS.

The column “Range/Parameter” shows either the adjust- ment ranges of the corresponding functions or a list of the corresponding options.

The column “Resolution” indicates the smallest steps that you can use to increment/decrement the corre- sponding values.

Function

Code

Range/ Parameter

Resolution

 

 

 

 

EVALUAT.

EV

0: TCG

1

MODE

 

1: DGS

 

 

 

 

 

PROBE-#

PB

1 - 16

1

 

 

 

 

P-NAME

PN

8 letters

 

 

 

 

 

DEL-VEL

DV

1000 - 9999 m/s

1

 

 

 

 

X-DIAM

XD

3 - 35 mm

0.1

 

 

 

 

X-FREQU

XF

0.5 - 10 MHz

0.25

 

 

 

 

REF-ECHO

RE

0: DSR

1

 

 

1: SDH

 

 

 

2: BW

 

 

 

 

 

REF-SIZE

RS

0: INFINITE

0.1

 

 

0.5 - 35 mm

 

 

 

 

 

DGS-30

Issue 02, 03/99

Krautkramer USN 52R/USN 52L Option DGS