Operation

 

Function check

 

 

 

 

 

 

Check

Result

Remark

 

 

 

4. Range

 

 

 

 

 

For settings see 2.

Initial pulse+Echo for RANGE/with Div.:

Up to RANGE = 25 mm only initial

RANGE: switch through from 5 mm to

50 mm / 0, 10

pulse and post-pulse oscillation

5000 mm

100 mm / 0, 5, 10

visible.

 

250 mm / 5th Echo at 10

Set RANGE to fixed step range

 

500 mm / 5th Echo at 5

 

1000 mm / 4th Echo at 2

and press M.

 

5000 mm / Echo sequence between 0 and 1

 

 

 

 

5. Pulse shift

 

 

 

 

 

See 2. for settings.

1st echo at Div. 4

Maximum shift is dependent on

DELAY approx. -17 µs

 

the range and the sound velocity.

 

 

 

DELAY approx. +17 µs

2nd echo at Div. 2

 

 

 

 

6. S/N ratio / Spurious oscillationg

 

 

 

 

 

For settings see 2. Also:

Both LEDs flash alternately.

 

PULSER: SINGLE

 

 

GATE LOGIC: POSITIVE

 

 

a-THRESH: 40%

 

 

a-START: 40 mm

 

 

a-WIDTH: 50 mm

 

 

GAIN so that 1st echo is just over the

 

 

gate

 

 

 

 

 

Press dB REF, then remove N30 and

read dB REF value: Db REF should be 80

Oscillations must not occur.

increase GAIN until the average value of

dB.

 

the noise is 40% SH

 

 

 

 

 

5-66

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

Page 111
Image 111