Echo evaluation with DAC/TCG

Operation

 

 

5.6Echo evaluation with

DAC/TCG

To be able to evaluate a flaw indication by means of the DAC/TCG, certain requirements have to be met:

The Distance-Amplitude Curve must already be recorded.

It only applies to the same probe that was used for recording the curve. Not even another probe of the same type must be used!

The curve and the DAC/TCG only apply to the material that corresponds with the material of the reference block.

All functions affecting the echo amplitude must be set as they were during recording of the curve. This especially applies to the following functions:

DAMPING, FREQ.,RECTIF., MTL VEL, PULSER and REJECT.

TOF is set to PEAK.

AMPLITUDE is set to dB THRESHOLD.

Choose the echo amplitude (as dB difference to the curve) using the TV -keys for an enlarged representation in the A-scan.

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-31

Page 76
Image 76
Epson USN 52L Echo evaluation with, DAMPING, FREQ.,RECTIF., MTL VEL, Pulser and Reject, Amplitude is set to dB Threshold