Storing measured values - Data Logger

Operation

 

 

Clearing FILES

All cleared data are irrevocably lost. Therefore, transfer the data to a PC or printer beforehand if necessary.

Operation:

Change to the second operating level and select the function group for the Data Logger.

Select the data set in FILE NUMBER which you wish to delete. Select ALL CLEAR to completely delete the Data Logger.

AAttention:

All deleted data are irretrievably lost!

Simultaneously press Min FILE NUMBER and keep it pressed for about 3 seconds.

When the FILE status “EMPTY” or “ALL CLEAR” is displayed, the selected FILE or the complete contents of the Data Logger are cleared.

Storing measured values

A measured value is only stored if it is not equal to 0 and if the active measurement location is empty. To start with, set the following values:

FILE NUMBER:

EMPTY or ACTIVE

FILE MODE:

THICKNESS

COPY MODE:

DATALOGGER

MEASURE:

0 TO 1st or MULTECHO

The thickness value is displayed in the A-scan mode at the top right-hand side. The measurement line shows:

LOC: the memory slot number of the selected FILE. The next measured value will be stored here.

VAL: the value which is stored on the displayed measurement location.

Operation:

Press A to store the measured value.

HNote:

You should transfer data to a PC in order to save them permanently. You can order the software which is re- quired for data transfer from Krautkrämer (refer to Chap- ter 2).

Krautkramer USN 52R/USN 52L

Issue 05, 02/00

5-37