Preparing the DGS evaluation

Operating the DGS

For the reference echo you need a test block with a known reflector. This reflector can be one of the follow- ing types:

backwall

disk-shaped reflector

side-drilled hole. With known diameter of the side- drilled hole, the diameter of a corresponding disk- shaped reflector can be calculated, provided that the

following requirements are met:

DSDH ³ 1,5 l s ³ 1,5 N

with DSDH:

diameter of side-drilled hole

s:

sound path in the test object

l:

wavelength in the test material

N:

near-field length of the probe used

 

in the test material

DESR:

diameter of the equivalent

 

disk-shaped reflector

 

 

 

 

 

 

 

DESR =

 

Ö

2

 

 

 

 

 

 

 

 

 

Ö p lÖsDSDH

Operation

HNote:

Before measuring, use M to select the setting PEAK in MEASURE. In addition, activate the gate.

Connect the required probe to the USN 52 R and select a suitable reference block for receiving the reference echo.

Adjust the A-scan and digital display (please refer to USN 52R/USN 52L Operating Manual, Chapter 5).

AAttention:

Proper A-scan and digital calibration will

assure that the MTL VEL and ZERO parameters are set correctly, which is necessary for an accurage DGS curve.

Switch to the second operation level: X

Tag the function table DGS: T or V

Tag the left column of the DGS function group using J or C.

Select the required probe number.

Krautkramer USN 52R/USN 52L Option DGS

Issue 02, 03/99

DGS-21

Page 196
Image 196