DGS mode

 

 

Overview of the DGS menu

 

 

 

 

Function

Description

Ranges/Options

Keys

X-FREQU

Frequency of the probe element

0.5 - 10 MHz for probe

N - decrement

 

 

nos. 1 - 13 are

O - increment

 

 

preallocated

 

 

 

 

 

REF-ECHO

Type of the reference reflector

DSR - (disk-shaped

N or O - indicates the required

 

used

reflector)

option

 

 

SDH - (side-drilled hole)

 

 

 

BW - (backwall)

 

 

 

 

 

REF-SIZE

Diameter of the reference reflector

INFINITE 0.5 - 35 mm

N - decrement

 

is automatically INFINITE if REF-

(limited by the probe

O - increment

 

ECHO is set to BW (backwall).

parameters)

 

 

 

 

 

ATT-REF

Sound attenuation in the reference

0 - 100 dB/m

N - decrement

 

block

 

O - increment

 

 

 

 

ATT-TEST

Sound attenuation in the test

0 - 100 dB/m

N - decrement

 

object

 

O - increment

 

 

 

 

AMPL-COR

Value for the amplitude correction.

-20 - +20 dB

N - decrement

 

This is required if you are using an

 

O - increment

 

angle-beam probe (value as per

 

 

 

probe data sheet)

 

 

 

 

 

 

T-LOSS

Transfer loss. Compensates for

-20 - +20 dB

N - decrement

 

differences in coupling conditions

 

O - increment

 

between reference and test block.

 

 

 

 

 

 

Krautkramer USN 52R/USN 52L Option DGS