Operation

Adjustment of the USN 52R/USN 52L

 

 

Clearing all data

You can clear all data sets (data logger and active in- strument settings).

Operation:

Switch off the instrument.

Keep keys F , E and K simultaneously for approximately 10 seconds.

All data sets are now cleared.

Displaying stored A-Scans (preview)

In the function group MEM you have the possibility to display all A-Scans belonging to the stored data sets without having to load them simultaneously. In this way, you obtain a quick overlook of your stored data.

Operation:

Simultaneously press N and O in the function SET#.

The selected data set’s A-Scan is displayed. At the same time, the function GAIN is changed in DATA SET PREVIEW as recognition of this preview function.

If additional information was stored for this data set by means of the MEMO function, the first 46 characters of this additional information are displayed in the menu bar (please refer to chapter 5.10).

You can scroll through all stored data sets using M. The corresponding A-Scans are displayed.

If you require to load a displayed data set, set the RECALL function to ON using Mand acknowledge system safety prompt with F. The preview function will then be switched off.

If you do not wish to load a data set: exit the preview function by pressing N and O simultaneously in SET#.

5-16

Issue 05, 02/00

Krautkramer USN 52R/USN 52L

Page 61
Image 61
Epson USN 52R, USN 52L manual Clearing all data, Displaying stored A-Scans preview