pH ARC SensorsModbus RTU Programmers Manual (EPHUM011)
624300/01 page 22/ 67
2.5.3 Primary Measurement Channel 6 (Temperature)
2.5.3.1 Description of PMC6
In register 2400, a plain text ASCII description of PMC6 is given
Start
register Number of
registers Reg1Reg8
16 ASCII characters Modbus
function code Read
access Write
access
2400 8 Description of PMC6 3,4 U/A/S none
Figure 2.5.3.1.1:Definition of register 2400.
Command: PMC6t ext Modbusaddress: 2400 Length: 8Type: 3Read
Parameter: Text
Format: Character
Value: T
Figure 2.5.3.1.2: Examplet o read the description. It is T(Temperature).
2.5.3.2 Selecting the Physical Unit for PMC6
In register 2408,the available physical units of PMC6 are defined.
Start
register Number of
registers Reg1/ Reg2
(bitwise defined) Modbus
function code Read
access Write
access
2408 2 Available physicalunits of PMC6 3, 4 U/A/S none
Figure 2.5.3.2.1:Definition of register 2408.
Command: PMC6available units Modbusaddress: 2408 Length: 2Type: 3Read
Parameter: Units
Format: Hex
Value: 0x06
Figure 2.5.3.2.2: Examplet o read the available physical unit for PMC6. K (0x02) + °C (0x04), total
0x06.
In register 2410,the active physical unit of PMC6 can be selected, by choosing one of the physical
units that are defined in register 2408.
Start
register Number of
registers Reg1/ Reg2
(bitwise defined) Modbus
function code Read
access Write
access
2410 2 Selected activephysical unit of PMC6 16 none U/A/S
Figure 2.5.3.2.3:Definition of register 2410. Only one bit can be set.
Command: PMC6 setunit Modbusaddress: 2410 Length: 2Type: 16 Write
Parameter: Unit
Format: Hex
Value: 0x04
Figure 2.5.3.2.4:Example to set the physical unit of PMC6 to °C (0x04).
Attention:
Changingt hephysical unit has also an influence on the output of AO1 / AO2, as the same physical unit
is activef or the analogoutputs. All limits of the 4-20 mA analog output have to be redefined after
changing the physicalunit!