pH ARC SensorsModbus RTU Programmers Manual (EPHUM011)
624300/01 page 20/ 67
2.5.2 Primary Measurement Channel 1 (pH)
2.5.2.1 Descriptionof PMC1
In register 2080, a plain text ASCII description of PMC1 is given.
Start
register Number of
registers Reg1Reg8
16 ASCII characters Modbus
function code Read
access Write
access
2080 8 Description ofPMC1 3, 4 U/A/S none
Figure 2.5.2.1.1:Definition of register 2080.
Command: PMC 1 text Modbusaddress: 2080 Length: 8Type: 3Read
Parameter: Text
Format: Character
Value: pH
Figure 2.5.2.1.2: Examplet o read the description. It is pH.
2.5.2.2 Selecting the Physical Unit for PMC1
In register 2088,the available physical units for this channel are defined.
Start
register Number of
registers Reg1/ Reg2
(bitwise defined) Modbus
function code Read
access Write
access
2088 2 Available physicalunits of PMC1 3, 4 U/A/S none
Figure 2.5.2.2.1:Definition of register 2088.
Command: PMC1available units Modbusaddress: 2088 Length: 2Type: 3Read
Parameter: Units
Format: Hex
Value: 0x201000
Figure 2.5.2.2.2: Example to read the available physical units of PMC1: pH (0x001000) + mV
(0x200000),total 0x201000.
In register 2090,the active physical unit for this channel can be selected, by choosing one of the
physical units that are defined in register 2088.
Start
register Number of
registers Reg1/ Reg2
(bitwise defined) Modbus
function code Read
access Write
access
2090 2 Selected activephysical unit for the
PMC1 16 none S
Figure 2.5.2.2.3:Definition of register 2090. Only one bit can be set.
Command: PMC1 set unit Modbusaddress: 2090 Length: 2Type: 16 Write
Parameter: Unit
Format: Hex
Value: 0x1000
Figure 2.5.2.2.4:Example to set the physical unit of PMC1 to pH (0x1000).
Attention:
Changingt hephysical unit has also an influence on the output of AO1 / AO2, as the same physical unit
is activef or the analogoutputs. All limits of the 4-20 mA analog output have to be redefined after
changing the physicalunit!