pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

The pH ARC Sensor register structure uses the following physical units used for Primary or Secondary Measurement Channels.

Bit #

Hex code

Physical unit

Start register. (8 ASCII characters, length 4 registers,

 

 

 

Type 3, read for U/A/S)

0 (LSB)

0x00000001

none

1920

1

0x00000002

K

1924

2

0x00000004

°C

1928

3

0x00000008

°F

1932

4

0x00000010

%-vol

1936

5

0x00000020

%-sat

1940

6

0x00000040

ug/l ppb

1944

7

0x00000080

mg/l ppm

1948

8

0x00000100

g/l

1952

9

0x00000200

uS/cm

1956

10

0x00000400

mS/cm

1960

11

0x00000800

1/cm

1964

12

0x00001000

pH

1968

13

0x00002000

mV/pH

1972

14

0x00004000

kOhm

1976

15

0x00008000

MOhm

1980

16

0x00010000

pA

1984

17

0x00020000

nA

1988

18

0x00040000

uA

1992

19

0x00080000

mA

1996

20

0x00100000

uV

2000

21

0x00200000

mV

2004

22

0x00400000

V

2008

23

0x00800000

mbar

2012

24

0x01000000

Pa

2016

25

0x02000000

Ohm

2020

26

0x04000000

%/°C

2024

27

0x08000000

°

2028

28

0x10000000

not used

2032

29

0x20000000

not used

2036

30

0x40000000

not used

2040

31 (MSB)

0x80000000

SPECIAL

2044

Figure 2.5.1.4: Definition of physical units used for PMC and SMC.

Command: Unit text

Modbus address: 1968

Length: 4

Type: 3

Read

Parameter:

Text

 

 

 

 

Format:

Character

 

 

 

 

Value:

pH

 

 

 

 

Figure 2.5.1.5: Example to read the physical unit in plain text ASCII in register 1968

624300/01

page 19 / 67

Page 19
Image 19
Hamilton Electronics EPHUM011 manual Special