pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.6.2.3Reading all Values for PA9

By reading register 3370, the active physical unit of measurement, the selected value, and the min and max values can be read.

Start

Number of

Reg1 /

 

Reg3 /

Reg5 /

 

Reg7 /

Modbus

Read

Write

register

registers

Reg2

 

Reg4

Reg6

 

Reg8

function

access

access

 

 

 

 

 

 

 

 

 

 

code

 

 

 

 

3370

8

 

Physical

 

Current

Min

 

Max

3, 4

 

 

U/A/S

none

 

 

 

unit

 

value

value

 

value

 

 

 

 

 

 

Figure 2.6.2.3.1: Definition of register 3370.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command: Moving average

 

Modbus address: 3370

Length: 8

 

Type: 3

Read

 

Parameter:

Unit

 

Value

 

Min value

 

 

Max value

 

 

Format:

 

Hex

 

Decimal

 

Decimal

 

 

Decimal

 

 

Value:

 

0x01

 

10

 

1

 

 

 

 

16

 

 

 

Figure 2.6.2.3.2: Example to read PA9. The physical unit is 0x01 (“none”), the value is 10 and the limit is 1 to 16.

624300/01

page 29 / 67

Page 29