pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7.5 Currently active Calibration Parameters part 1

In registers 5164 (CP1), 5196 (CP2) and 5324 (CP6) the currently active calibration parameters part 1 are stored. These registers contain the values for temperature, number of calibrations and operating hour upon calibration.

Start

Number

Reg1 / Reg2

Reg3 /

Reg5 /

Reg7 /

Modbus

Read

Write

register

of

 

Reg4

Reg6

Reg8

function

access

access

 

registers

 

 

 

 

code

 

 

5164

8

Unit of

Value of

Number of

Operating

3, 4

U/A/S

none

 

 

temperature for

temperature

calibrations

hour for

 

 

 

 

 

CP1 (bitwise

of CP1

at CP1

CP1

 

 

 

 

 

defined)

 

 

 

 

 

 

5196

8

Unit of

Value of

Number of

Operating

3, 4

U/A/S

none

 

 

temperature for

temperature

calibrations

hour for

 

 

 

 

 

CP2 (bitwise

of CP2

at CP2

CP2

 

 

 

 

 

defined)

 

 

 

 

 

 

5324

8

Unit of

Value of

Number of

Operating

3, 4

U/A/S

none

 

 

temperature for

temperature

calibrations

hour for

 

 

 

 

 

CP6 (bitwise

of CP6

at CP6

CP6

 

 

 

 

 

defined)

 

 

 

 

 

 

Figure 2.7.5.1: Definition of register 5164 for CP1, 5196 for CP2 and 5324 for CP6.

Command: Calibration CP1 values

Modbus address: 5164

 

Length: 8

Type: 3

Read

Parameter:

Unit of temperature

Temperature

Number

of cali

Operating hour

Format:

Hex

Float

Decimal

 

Float

 

Value:

0x00000004

24.35184

6

 

 

23.78

 

Figure 2.7.5.2: Example to read the calibration values for CP1. The physical unit is °C (0x00000004), the temperature is 24.35184 °C, the number of calibrations at CP1 is 6 and the operating hour is 23.78 h.

Command: Calibration CP2 values

Modbus address: 5196

 

Length: 8

Type: 3

Read

Parameter:

Unit of temperature

Temperature

Number

of cali

Operating hour

Format:

Hex

Float

Decimal

 

Float

 

Value:

0x00000004

24.37691

5

 

 

16.45

 

Figure 2.7.5.3: Example to read the calibration values for CP2. The physical unit is °C (0x00000004), the temperature is 24.37691 °C, the number of calibrations at CP1 is 5 and the operating hour is 16.45 h.

Command: Calibration CP6 values

Modbus address: 5324

 

Length: 8

Type: 3

Read

Parameter:

Unit of temperature

Temperature

Number

of cali

Operating hour

Format:

Hex

Float

Decimal

 

Float

 

Value:

0x00000004

29.93368

12

 

 

379.5167

 

Figure 2.7.5.4: Example to read the calibration values 1 for CP6. The physical unit is °C (4), the temperature is 29.93 (°C), the number of calibrations at CP1 is 12 and the operating hour is

379.51(h).

2.7.6Currently active Calibration Parameters part 2

Registers 5172 (CP1), 5204 (CP2) and 5332 (CP6) are not defined for pH ARC Sensors, as they document atmospheric pressure and salinity used for VISIFERM DO / VISIFERM DO ARC Sensors only.

624300/01

page 48 / 67

Page 48
Image 48
Hamilton Electronics EPHUM011 manual Currently active Calibration Parameters part, 0x00000004 24.35184 23.78