Hamilton Electronics EPHUM011 manual Definition of Errors

Models: EPHUM011

1 67
Download 67 pages 49.01 Kb
Page 61
Image 61

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.8.4.3Definition of Errors

Bit #

Hex

Description

 

 

not available

5

0x0000020

Glass resistance too high

6

0x0000040

Glass resistance too low

7

0x0000080

Reference electrode resistance too high

8

0x0000100

Reference electrode resistance too low

 

 

not available

15

0x0008000

Auxiliary electrode electrical potential too high

16

0x0010000

Auxiliary electrode electrical potential too low

17

0x0020000

Auxiliary electrode resistance too high

18

0x0040000

Auxiliary electrode resistance too low

 

 

not available

250x2000000 Temperature sensor defective Figure 2.8.4.3.1: Definition of errors “measurement”.

Bit #

Hex

Description

10x0000002 Sensor failure (Quality value < 15%) Figure 2.8.4.3.2: Definition of errors “calibration”.

Bit #

Hex

Description

 

 

not available

Figure 2.8.4.3.3: Definition of errors “interface”. None is defined.

Bit #

Hex

Description

 

 

not available

240x1000000 Internal communication error (between front-end and user-end) Figure 2.8.4.3.4: Definition of errors “hardware”.